How To Use USB-DAQ Memphis TN

Our two previous columns discussed USB-DAQ software, timing and triggering. This column wraps up th...

Local Companies

Daco Trailer Leasing
(901) 332-0700
1761 E Brooks Rd
Memphis, TN
Rentmax
(901) 398-7368
4080 Elvis Presley Blvd
Memphis, TN
Rent-A-Center
(901) 353-5300
2570 Frayser Blvd
Memphis, TN
Kenesaw Leasing Inc
(901) 767-9670
665 Oakleaf Office Ln
Memphis, TN
Rent-A-Center
(901) 323-4799
3911 Park Ave
Memphis, TN
Future Rental Services
(901) 333-7368
2205 Lamar Ave
Memphis, TN
Rent-A-Center
(901) 377-8300
4005 Jackson Ave
Memphis, TN
Rent-A-Center
(901) 396-1400
4090 Elvis Presley Blvd
Memphis, TN
Enterprise Rental Car
(901) 624-8377
7420 Winchester Rd
Memphis, TN
Enterprise Financial Solutions Inc
(901) 624-3773
6750 Poplar Pike
Memphis, TN

provided by: Design News

Our two previous columns discussed USB-DAQ software, timing and triggering. This column wraps up the USB topic with a few aspects of USB connections engineers might not think about.

Most USB connectors provide two signal contacts, a ground and a +5V connection. That means your DAQ equipment shares a common ground with a host PC and other USB-based equipment, such as other DAQ or control devices, a printer, scanner and so on. Unfortunately, in most experiments or tests, the last thing you want is this common-ground connection, which can lead to ground loops. Past columns have covered grounding and ground loops, so there's no need to provide the details here. (See Useful Links, below). Search Google for USB and ?ground loop? and you'll find many cries for help.

You cannot simply snip the ground signal in a USB cable to ?break? a ground loop because the USB data signals need a ground reference. Thus, isolation must occur within your DAQ equipment between its I/O lines and an internal USB transceiver. (Data sheets should include isolation spec). If you suspect ground-loop problems in installed equipment, you can buy USB isolators and some USB hubs also offer electrical isolation. In addition to preventing ground-loop problems, isolated DAQ equipment protects your computer and other devices from accidental short circuits and high voltages at a USB-DAQ module. Suppose a 120V ac power line touches a sensor input. Isolation devices keep that ruinous potential from reaching your PC.

If you plan to use USB-DAQ modules, consider buying units that isolate your computer's USB ground from the local ground at your DAQ equipment. That means you should not try to power DAQ modules from the USB's +5V supply and you ? or your DAQ-equipment vendor ? must provide a separate power supply. Ensure that supply does not create its own ground loop!

In some cases, you must use +5V power from a USB port. In theory, you can draw up to 500 mA from a USB port, but you shouldn't try to draw more than 100 mA from a battery-powered host laptop. You may have seen instructions on USB equipment that caution against connecting it to USB-powered hubs. That's because USB cables and devices experience voltage drops and hubs use power. The official USB 2.0 specifications include a voltage-drop budget (see figure, above). The farther you go from your actual power source ? a host PC or powered hub ? the greater the possible voltage drop. In some cases, you can measure just over 4V at the end of a long USB run through a hub. Voltage drops can exceed those in the USB spec and affect DAQ measurements. Transients also can affect remote voltages and ground potentials can vary, too. When possible, use local power with USB-DAQ devices.

Useful Links

  1. ?Subtle Traps Degrade Sensors Signals,? http://rbi.ims.ca/5716-523.
  2. ?Isolation Technologies for Reliable Industrial Measurements,? http://rbi.ims.ca/5716-524.
  3. To build a ground-loop detector, see: http://rbi.ims.ca/5716-525.

Jon Titus, a former designer and chief editor of EDN and Test & Measurement World magazines, remembers when ?fast? signals operated at 10 MHz and programs came on paper tape.



author: By Jon Titus, Contributing Writer

Design News. Copyright © 2008 Reed Business Information, a division of Reed Elsevier Inc. All rights reserved.